Features
- Non-contact, laser measurement does not deform or mark measured objects.
- Available in 27 or 57mm maximum measurement models.
- Provides statistical analysis of measurements.
- Available fitted with optional rotary chuck or V-groove sample holders.
- Extremely easy to use.
- For off-line, stand-alone use in incoming inspection, quality control and laboratory applications.
Introduction
Derived from the highly successful InteliSENS® DG Series of scanning-laser diameter gauges, the InteliSENS® BG Series bench gauges are intended for off-line, stand-alone use in incoming inspection, quality control and laboratory applications.
A Proton Products InteliSENS® DG Series scanning-laser diameter gauge is integrated together with a miniature touch screen PC and a mains power supply unit on an aluminium baseboard. Depending on the application, optional ground steel V-grooves or a rotary chuck may also be integrated for precise sample holding.
InteliSENS® BG Series bench gauges are a fast, easy-to-use and non-contact alternative to vernier callipers and micrometer screw gauges. The touch screen PC provides measurement display with automatic maximum/minimum and ovality calculation, statistical analysis and data logging. Logged data may be extracted from the touch screen PC using a USB flash memory stick.
Operating Principle
Proton Products InteliSENS® DG Series diameter gauges rapidly scan a laser beam across the object under measurement. The laser beam is focused by a receiving lens onto a photodiode and the electrical signal digitally processed by a Proton-designed FPGA processor. The time duration for which the object obscures the laser beam is directly proportional to the object diameter.
Specifications
Model | BG1030 | BG1060 | Units |
---|---|---|---|
Gate size | 30 | 60 | mm |
1.18 | 2.36 | inch | |
Minimum diameter | 0.1 | 0.2 | mm |
0.004 | 0.004 | inch | |
Maximum diameter | 27 | 57 | mm |
1.063 | 2.244 | inch | |
Resolution | 1 | 1 | µm |
Accuracy | ±(1µm + 0.008% of object diameter) | ±(1.5µm + 0.008% of object diameter) | – |
Specification | Minimum | Typical | Maximum | Units |
---|---|---|---|---|
Measurement rate | 600 | scan/s | ||
Operating temperature | 5 | 45 | °C | |
41 | 113 | °F | ||
Relative humidity (at 38°C / 105°F) | 90 | % | ||
Power supply voltage | 85 | 220 | 264 | VAC (rms) |
Power supply frequency | 47 | 65 | Hz | |
Power consumption | 24 | W |