InteliSENS® BG Series

Quote

InteliSENS® BG Series

Non-Contact Diameter Measurement

Derived from the highly successful InteliSENS® DG Series of scanning-laser diameter gauges, the InteliSENS® BG Series bench gauges are intended for off-line, stand-alone use in incoming inspection, quality control and laboratory applications.

A Proton Products InteliSENS® DG Series scanning-laser diameter gauge is integrated together with a miniature touch screen PC and a mains power supply unit on an aluminium baseboard. Depending on the application, optional ground steel V-grooves or a rotary chuck may also be integrated for precise sample holding.

Features

  • Non-contact, laser measurement does not deform or mark measured objects.
  • Available in 27 or 57mm maximum measurement models.
  • Provides statistical analysis of measurements.
  • Available fitted with optional rotary chuck or V-groove sample holders.
  • Extremely easy to use.
  • For off-line, stand-alone use in incoming inspection, quality control and laboratory applications.

Introduction

Derived from the highly successful InteliSENS® DG Series of scanning-laser diameter gauges, the InteliSENS® BG Series bench gauges are intended for off-line, stand-alone use in incoming inspection, quality control and laboratory applications.

A Proton Products InteliSENS® DG Series scanning-laser diameter gauge is integrated together with a miniature touch screen PC and a mains power supply unit on an aluminium baseboard. Depending on the application, optional ground steel V-grooves or a rotary chuck may also be integrated for precise sample holding.

InteliSENS® BG Series bench gauges are a fast, easy-to-use and non-contact alternative to vernier callipers and micrometer screw gauges. The touch screen PC provides measurement display with automatic maximum/minimum and ovality calculation, statistical analysis and data logging. Logged data may be extracted from the touch screen PC using a USB flash memory stick.

Operating Principle

Proton Products InteliSENS® DG Series diameter gauges rapidly scan a laser beam across the object under measurement. The laser beam is focused by a receiving lens onto a photodiode and the electrical signal digitally processed by a Proton-designed FPGA processor. The time duration for which the object obscures the laser beam is directly proportional to the object diameter.

DG Operating Principle

Specifications

Model BG1030 BG1060 Units
Gate size 30 60 mm
1.18 2.36 inch
Minimum diameter 0.1 0.2 mm
0.004 0.004 inch
Maximum diameter 27 57 mm
1.063 2.244 inch
Resolution 1 1 µm
Accuracy ±(1µm + 0.008% of object diameter) ±(1.5µm + 0.008% of object diameter)
Specification Minimum Typical Maximum Units
Measurement rate 600 scan/s
Operating temperature 5 45 °C
41 113 °F
Relative humidity (at 38°C / 105°F) 90 %
Power supply voltage 85 220 264 VAC (rms)
Power supply frequency 47 65 Hz
Power consumption 24 W